Sensing Application — Ag doped Co/Ni bimetallic organic framework for determination of luteolin

Measurement evidence

Sensing Application

Ag doped Co/Ni bimetallic organic framework for determination of luteolin · Tang J., Hu T., Li N. et al. · Microchemical Journal · 2022 · 107461

5 measurement groups · 20 results

Reported values remain attached to the sample, method, conditions, extraction quality and source location that produced them.

differential pulse voltammetry (DPV) calibration

Ag-CoNi-MOF/GCE · Electrode

Luteolin concentrations 2.0 nM, 4.0 nM, 10.0 nM, 20.0 nM, 50.0 nM, 0.1 uM, 0.2 uM, 0.4 uM, 0.6 uM, 0.8 uM and 1.0 uM in 0.1 M BR buffer, pH 3.0.

Geometry
Ag-CoNi-MOF/GCE
Context
composite_sample
Measurement source
6 · 3.6 · Fig. 4; Table 1
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
luteolin DPV calibration intercept0.041 uAText
Exact Reported
6-7 · 3.6; 3.8 · Fig. 4b
luteolin DPV calibration R2R2 = 0.992Text
Exact Reported
6 · 3.6 · Fig. 4b
luteolin DPV calibration slopeMarked as a best value within this paperIpa (uA) = 15.71 C (uM) + 0.041Text
Exact Reported
6 · 3.6 · Fig. 4b; Table 1
luteolin DPV linear rangeMarked as a best value within this paper0.002-1.0 uMText
Range
1,6 · Abstract; 3.6 · Fig. 4; Table 1
luteolin detection limitMarked as a best value within this paper0.4 nM (S/N = 3)0.0004 uMS/N = 3Text
Exact Reported
1,6 · Abstract; 3.6 · Table 1

amperometric i-t interference/selectivity test

Ag-CoNi-MOF/GCE · Electrode

0.5 uM luteolin; 5-fold glucose, cysteine, citric acid, dopamine, ascorbic acid and quercetin; 50-fold K+, Na+, Mg2+, Ca2+, Fe3+.

Geometry
Ag-CoNi-MOF/GCE
Context
composite_sample
Measurement source
3 · 3.7 · Fig. S3c
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
ion interferent concentration multiple50-fold concentration of K+, Na+, Mg2+, Ca2+, Fe3+Text
Exact Reported
7 · 3.7 · Fig. S3c
organic interferent concentration multiple5-fold concentration of glucose, cysteine, citric acid, dopamine and ascorbic acid; 5-fold quercetinText
Exact Reported
7 · 3.7 · Fig. S3c
interference outcomeMarked as a best value within this paperadditives did not remarkably affect amperometric current response of luteolinQualitative
Qualitative
7 · 3.7 · Fig. S3c
luteolin amperometric current step visual estimatecurrent increases from approximately 0.5 uA baseline to approximately 0.8 uA after 0.5 uM luteolin additionvisual read from I-t traceVisual Estimate
Approximate
3 · Supporting information · Fig. S3c

DPV reproducibility

Ag-CoNi-MOF/GCE · Electrode

Six different Ag-CoNi-MOF sensors used to determine 1.0 uM luteolin by DPV.

Geometry
Ag-CoNi-MOF/GCE
Context
composite_sample
Measurement source
3 · 3.7 · Fig. S3a
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
reproducibility RSD for six sensors3.21%n = 6 sensorsText
Exact Reported
7 · 3.7 · Fig. S3a
six-electrode DPV current visual rangeapproximately 14.0-15.2 uA across six electrodesvisual range from bar chartVisual Estimate
Range
3 · Supporting information · Fig. S3a

intermittent DPV stability test

Ag-CoNi-MOF/GCE · Electrode

Current for 1.0 uM luteolin assessed every 2 days over 20 days.

Geometry
Ag-CoNi-MOF/GCE
Context
composite_sample
Measurement source
3 · 3.7 · Fig. S3b
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
20-day stability current visual estimateapproximately 14.0 uA at 20 daysvisual read from bar chartVisual Estimate
Approximate
3 · Supporting information · Fig. S3b
retained current after 20 daysMarked as a best value within this paper96.2% of initial current after 20 days20 daysText
Exact Reported
7 · 3.7 · Fig. S3b

DPV standard addition in human urine

Ag-CoNi-MOF/GCE · Electrode

Human urine diluted 10-fold with BR buffer pH 3.0; luteolin added at 0, 50, 100 and 500 nM; n = 4.

Geometry
Ag-CoNi-MOF/GCE
Context
composite_sample
Measurement source
7 · 3.8 · Table 2
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
human urine recovery rangeMarked as a best value within this paper98.3-103.0%RSD values below 4%Text
Range
7 · 3.8 · Table 2
urine sample 2 found concentration after 50 nM addition51.5 +/- 0.9 nM+/- 0.9 nMTable
Exact Reported
7 · 3.8 · Table 2
urine sample 2 recovery after 50 nM addition103.0%RSD 1.73%Table
Exact Reported
7 · 3.8 · Table 2
urine sample 3 found concentration after 100 nM addition98.3 +/- 2.8 nM+/- 2.8 nMTable
Exact Reported
7 · 3.8 · Table 2
urine sample 3 recovery after 100 nM addition98.3%RSD 2.85%Table
Exact Reported
7 · 3.8 · Table 2
urine sample 4 found concentration after 500 nM addition506 +/- 16.2 nM+/- 16.2 nMTable
Exact Reported
7 · 3.8 · Table 2
urine sample 4 recovery after 500 nM addition101.2%RSD 3.20%Table
Exact Reported
7 · 3.8 · Table 2