Sensing Application — Glucose sensing performance of bimetallic MOFs CoFe-ZIF/CC for enzyme-free saliva sensor applications

Measurement evidence

Sensing Application

Glucose sensing performance of bimetallic MOFs CoFe-ZIF/CC for enzyme-free saliva sensor applications · Liu J., Shi L., Shi Y. et al. · Journal of Alloys and Compounds · 2024 · 174733

8 measurement groups · 30 results

Reported values remain attached to the sample, method, conditions, extraction quality and source location that produced them.

Chronoamperometry glucose calibration

Co0.95Fe0.05-ZIF/CC/GCE electrode · Electrode

CA response to successive glucose additions at 0.40 V in 0.10 M NaOH; GCE-based Co0.95Fe0.05-ZIF/CC sensor.

Temperature
room temperature
Geometry
Co0.95Fe0.05-ZIF/CC/GCE in three-electrode cell.
Context
Main composite electrode sensitivity and LOD.
Measurement source
6-7 · 3.2 Glucose-Sensing Performance · Fig. 5a-b; Table 1
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
GCE high-range calibration equation from Fig. 5by3 = 103.90x + 108.10plot annotationFigure Axis
Rounded Reported
7 · Figure · Fig. 5b
GCE low-range calibration equation from Fig. 5by1 = 129.10x + 6.85plot annotationFigure Axis
Rounded Reported
7 · Figure · Fig. 5b
GCE mid-range calibration equation from Fig. 5by2 = 228.36x + 15.68plot annotationFigure Axis
Rounded Reported
7 · Figure · Fig. 5b
GCE sensor glucose detection limitMarked as a best value within this paperabout 5 uM; abstract/conclusions state less than 5 uMabout/less thanText
Approximate
1,6 · Abstract; 3.2 Glucose-Sensing Performance
Best GCE sensor sensitivityMarked as a best value within this paper3232.30 uA mM-1 cm-2 in 0.20-1.00 mMText
Exact Reported
1,6 · Abstract; 3.2 Glucose-Sensing Performance · Fig. 5b; Table 1
GCE sensor sensitivity in high glucose range1470.63 uA mM-1 cm-2 in 1.00-2.00 mMText
Exact Reported
6 · 3.2 Glucose-Sensing Performance · Fig. 5b; Table 1
GCE sensor sensitivity in low glucose range1827.25 uA mM-1 cm-2 in 0.005-0.20 mMText
Exact Reported
6 · 3.2 Glucose-Sensing Performance · Fig. 5b; Table 1

Chronoamperometry glucose calibration

Co0.95Fe0.05-ZIF/GCE electrode · Electrode

Control CA response for Co0.95Fe0.05-ZIF at 0.40 V; values reported from Fig. S9a-e.

Temperature
room temperature
Geometry
Co0.95Fe0.05-ZIF/GCE in three-electrode cell.
Context
Pristine bimetallic ZIF control.
Measurement source
6 · 3.2 Glucose-Sensing Performance · Fig. S9a-e
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Pristine Co0.95Fe0.05-ZIF high-range calibration equation from SI Fig. S9y3 = 2.26E-2x + 23.36plot annotationFigure Axis
Rounded Reported
7 · Figure S9 · Fig. S9e
Pristine Co0.95Fe0.05-ZIF low-range calibration equation from SI Fig. S9y1 = 3.09E-2x + 2.87plot annotationFigure Axis
Rounded Reported
7 · Figure S9 · Fig. S9e
Pristine Co0.95Fe0.05-ZIF mid-range calibration equation from SI Fig. S9y2 = 3.81E-2x + 1.34plot annotationFigure Axis
Rounded Reported
7 · Figure S9 · Fig. S9e
Pristine Co0.95Fe0.05-ZIF sensitivity high range319.75 uA mM-1 cm-2 in 1.50-2.10 mMText
Exact Reported
6 · 3.2 Glucose-Sensing Performance · Fig. S9
Pristine Co0.95Fe0.05-ZIF sensitivity low range437.28 uA mM-1 cm-2 in 0.025-0.30 mMText
Exact Reported
6 · 3.2 Glucose-Sensing Performance · Fig. S9
Pristine Co0.95Fe0.05-ZIF sensitivity mid range538.60 uA mM-1 cm-2 in 0.30-1.50 mMText
Exact Reported
6 · 3.2 Glucose-Sensing Performance · Fig. S9

Chronoamperometry potential optimisation

Co0.95Fe0.05-ZIF/CC/GCE electrode · Electrode

Current response measured at 0.20, 0.30, 0.40 and 0.45 V for glucose additions; 0.40 V selected for subsequent CA.

Temperature
room temperature
Geometry
Three-electrode GCE cell.
Context
Composite electrode.
Measurement source
5-6 · 3.2 Glucose-Sensing Performance · Fig. 4e-f
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Appropriate operating-potential range0.20-0.40 VrangeText
Range
5 · 3.2 Glucose-Sensing Performance · Fig. 4a
Optimised operating potential for CAMarked as a best value within this paper0.40 VText
Exact Reported
5 · 3.2 Glucose-Sensing Performance · Fig. 4e-f

Noninvasive PET glucose sensor CA calibration

Co0.95Fe0.05-ZIF/CC/PET saliva sensor · Electrode

Screen-printed PET three-electrode sensor with Co0.95Fe0.05-ZIF/CC; 30 uL glucose solution dropped on working region; glucose solutions from 0.005 mM to 2.10 mM prepared.

Temperature
room temperature
Geometry
PET three-electrode system coated with Co0.95Fe0.05-ZIF/CC and KOH-PVA gel.
Context
Application-level saliva sensor device.
Measurement source
8-9 · 3.4 Glucose-sensing performance of non-enzymatic glucose sensor · Fig. 6b-c
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
PET sensor calibration equation from Fig. 6cy = 3.81E-2x + 1.52plot annotationFigure Axis
Rounded Reported
9 · Figure · Fig. 6c
PET sensor sensitivity539.56 uA mM-1 cm-2Text
Exact Reported
8 · 3.4 Glucose-sensing performance · Fig. 6c
PET sensor reported linear rangefrom 0.025 uM to 2.00 mMas printed; likely unit inconsistencyText
Uncertain
8 · 3.4 Glucose-sensing performance · Fig. 6c
PET sensor glucose detection limit25 uMText
Exact Reported
8 · 3.4 Glucose-sensing performance · Fig. 6c

PET sensor stability, repeatability and reproducibility

Co0.95Fe0.05-ZIF/CC/PET saliva sensor · Electrode

PET sensor retention after 30 days; RSD for 10 measurements on one PET sensor and for multiple devices.

Temperature
room temperature
Geometry
Co0.95Fe0.05-ZIF/CC/PET sensor.
Context
Application device durability statistics.
Measurement source
8-9 · 3.4 Glucose-sensing performance · Fig. 6d-f
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
PET sensor repeatability RSDless than 12.17% for 10 measurementsless thanText
Exact Reported
8 · 3.4 Glucose-sensing performance · Fig. 6e
PET sensor reproducibility RSD (figure annotation)RSD <= 18.20%less than or equalFigure Axis
Rounded Reported
9 · Figure · Fig. 6f
PET sensor reproducibility RSD (text)less than 17.70% for five devicesless thanText
Exact Reported
8 · 3.4 Glucose-sensing performance · Fig. 6f
PET sensor 30-day retention70.16% after 30 daysText
Exact Reported
8 · 3.4 Glucose-sensing performance · Fig. 6d

Saliva real-sample comparison with commercial blood glucose meter

Co0.95Fe0.05-ZIF/CC/PET saliva sensor · Electrode

Saliva from ten volunteers collected before meal and one hour after meal; blood glucose measured under same conditions for comparison.

Geometry
Co0.95Fe0.05-ZIF/CC saliva sensor with BIOSYS/mobile-phone readout.
Context
Real-sample application.
Measurement source
8,10 · 3.5 Real sample detection · Fig. 7
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Saliva sensor vs commercial blood glucose trendCurrent response after a meal increased distinctly and was consistent with commercial blood glucose meter trends.Text
Qualitative
8,10 · 3.5 Real sample detection · Fig. 7b
Real-sample volunteer countten volunteersText
Exact Reported
8 · 3.5 Real sample detection · Fig. 7b

CA selectivity/interference test

Co0.95Fe0.05-ZIF/CC/GCE electrode · Electrode

Successive addition of 500 uM glucose, urea, sucrose, L-serine, L-lysine, galactose, D-xylose, D-fructose and glucose.

Temperature
room temperature
Geometry
Co0.95Fe0.05-ZIF/CC/GCE.
Context
Composite electrode anti-interference performance.
Measurement source
7-8 · 3.3 Selectivity, stability, reproducibility and repeatability · Fig. 5c
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Selectivity interferent concentration500 uM glucose and 500 uM interferentsText
Exact Reported
7 · 3.3 Selectivity · Fig. 5c
Selectivity resultCurrent response increases greatly after glucose; negligible responses for urea, sucrose, L-serine and L-lysine, slight responses for galactose, D-xylose and D-fructose.Text
Qualitative
7 · 3.3 Selectivity · Fig. 5c

Long-term stability, repeatability and reproducibility tests

Co0.95Fe0.05-ZIF/CC/GCE electrode · Electrode

GCE-based composite sensor retention after 30 days; RSD for 10 repeated measurements on one electrode and for ten Co0.95Fe0.05-ZIF/CC/GCE samples.

Temperature
room temperature
Geometry
Co0.95Fe0.05-ZIF/CC/GCE.
Context
Composite electrode durability statistics.
Measurement source
7-8 · 3.3 Selectivity, stability, reproducibility and repeatability · Fig. 5d-f
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
GCE sensor repeatability RSDless than 6.50% for 10 measurements on one electrodeless thanText
Exact Reported
8 · 3.3 Selectivity, stability, reproducibility and repeatability · Fig. 5e
GCE sensor reproducibility RSDless than 6.43% for ten Co0.95Fe0.05-ZIF/CC/GCE samplesless thanText
Exact Reported
8 · 3.3 Selectivity, stability, reproducibility and repeatability · Fig. 5f
GCE sensor 30-day current-density retention76.10% after 30 daysText
Exact Reported
7-8 · 3.3 Selectivity, stability, reproducibility and repeatability · Fig. 5d