Sensing Application — An examination of some active organometallic substances for ion-selective electrodes

Measurement evidence

Sensing Application

An examination of some active organometallic substances for ion-selective electrodes · Sharp M. · Analytica Chimica Acta · 1975 · 165-176

9 measurement groups · 25 results

Reported values remain attached to the sample, method, conditions, extraction quality and source location that produced them.

Selectrode solid-state electrode potentiometric response

Cd-Fe(III) coordination polymer solid-state electrode · Electrode

Cd-Fe coordination polymer electrode response in cadmium nitrate and sodium nitrate, bromide, iodide, perchlorate and sulphate solutions

Temperature
298
Geometry
Selectrode solid-state electrode
Context
mixed-metal Cd-Fe coordination polymer active solid
Measurement source
p008-p009 / journal pages 172-173 · Results; Coordination polymers · Figs. 13-14
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Cd-Fe coordination polymer reproducibility+/- 3 mV+/-Text
Exact Reported
p009 / journal page 173 · Results; Coordination polymers
Cd-Fe coordination polymer response time5-10 minrangeText
Range
p009 / journal page 173 · Results; Coordination polymers
Cd-Fe coordination polymer response typeanion rather than cation response; nitrate response with slight anion selectivityQualitative
Qualitative
p009 / journal page 173 · Results; Coordination polymers · Figs. 13-14

Selectrode solid-state electrode potentiometric response

copper coordination polymer solid-state electrode · Electrode

Copper-containing coordination polymer electrode response in Cu(NO3)2 solutions

Temperature
298
Geometry
Selectrode solid-state electrode
Context
pristine copper coordination polymer active solid
Measurement source
p008-p009 / journal pages 172-173 · Results; Coordination polymers · Fig. 12
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
copper coordination polymer Cu2+ linear response range10^-1-10^-3 M Cu2+rangeText
Range
p009 / journal page 173 · Results; Coordination polymers · Fig. 12
copper coordination polymer response timeless than 2-3 min<; 2-3 rangeText
Range
p009 / journal page 173 · Results; Coordination polymers
copper coordination polymer selectivitysome selectivity for copper(II) over nickel(II) and alkali metal cationsQualitative
Qualitative
p009 / journal page 173 · Results; Coordination polymers
copper coordination polymer Cu2+ response slope33 mV decade-1Text
Rounded Reported
p009 / journal page 173 · Results; Coordination polymers · Fig. 12

Selectrode solid-state electrode potentiometric response

Cu-TCNE polymer solid-state electrode · Electrode

Cu-TCNE electrode response in Cu(NO3)2 solutions; selectivity checked against Ni2+, Zn2+, Mg2+, and K+

Temperature
298
Geometry
Selectrode solid-state electrode
Context
pristine Cu-TCNE polymer active solid
Measurement source
p007-p008 / journal pages 171-172 · Results; TCNE polymers · Fig. 10
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Cu-TCNE Cu2+ response rangeMarked as a best value within this paper10^-1-10^-5 M Cu2+rangeText
Range
p007 / journal page 171 · Results; TCNE polymers · Fig. 10
Cu-TCNE selectivityMarked as a best value within this paperselectivity towards Cu2+ over Ni2+, Zn2+, Mg2+ and K+Qualitative
Qualitative
p008 / journal page 172 · Results; TCNE polymers
Cu-TCNE Cu2+ response slopeMarked as a best value within this paper29 mV decade-1Text
Rounded Reported
p007 / journal page 171 · Results; TCNE polymers · Fig. 10

Selectrode solid-state electrode potentiometric response

MgPc solid-state electrode · Electrode

Measured in Mg(NO3)2 and sodium nitrate/chloride/sulphate solutions at 25 C

Temperature
298
Geometry
Selectrode solid-state electrode
Context
pristine MgPc active solid
Measurement source
p006-p007 / journal pages 170-171 · Results; Metal-phthalocyanines · Figs. 7-8
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
MgPc response typeanion rather than cation sensitivity; nitrate activity variations prevailQualitative
Qualitative
p007 / journal page 171 · Results; Metal-phthalocyanines · Figs. 7-8

Selectrode solid-state electrode potentiometric response

Mg-TCNE polymer solid-state electrode · Electrode

Mg-TCNE electrode response in Mg(NO3)2 solutions; selectivity checked against Ca2+, Na+, K+, Zn2+, Ni2+, Cu2+

Temperature
298
Geometry
Selectrode solid-state electrode
Context
pristine Mg-TCNE polymer active solid
Measurement source
p007 / journal page 171 · Results; TCNE polymers · Fig. 9
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Mg-TCNE Mg2+ linear response range10^-1-10^-4 M Mg2+rangeText
Range
p007 / journal page 171 · Results; TCNE polymers · Fig. 9
Mg-TCNE potential reproducibility+/- 2 mV+/-Text
Exact Reported
p007 / journal page 171 · Results; TCNE polymers
Mg-TCNE response timeMarked as a best value within this paper< 2 min<Text
Range
p007 / journal page 171 · Results; TCNE polymers
Mg-TCNE selectivity/interferencelittle or no selectivity for Mg2+ over Ca2+, Na+, K+; Zn, Ni, Cu strongly interferedQualitative
Qualitative
p007 / journal page 171 · Results; TCNE polymers
Mg-TCNE Mg2+ response slopeapproximately 33 mV decade-1approximatelyText
Approximate
p007 / journal page 171 · Results; TCNE polymers · Fig. 9

Selectrode solid-state electrode potentiometric response

Ni-TCNE polymer solid-state electrode · Electrode

Ni-TCNE and Zn-TCNE electrodes tested in Ni(NO3)2 and Zn(NO3)2 solutions respectively

Temperature
298
Geometry
Selectrode solid-state electrode
Context
pristine Ni-TCNE and Zn-TCNE polymer active solids
Measurement source
p008 / journal page 172 · Results; TCNE polymers · Fig. 11
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Ni-TCNE and Zn-TCNE response range10^-1-10^-3 M M2+rangeText
Range
p008 / journal page 172 · Results; TCNE polymers · Fig. 11
Ni-TCNE and Zn-TCNE slope behaviourslopes significantly less than Nernst relationQualitative
Qualitative
p008 / journal page 172 · Results; TCNE polymers · Fig. 11

Selectrode solid-state electrode potentiometric response

PbPc solid-state electrode · Electrode

PbPc, CuPc, and FePc solid-state electrodes tested in metal nitrate solutions

Temperature
298
Geometry
Selectrode solid-state electrode
Context
pristine phthalocyanine active solids
Measurement source
p007 / journal page 171 · Results; Metal-phthalocyanines
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
FePc response reproducibilityso irreproducible that no definite trends could be discernedQualitative
Qualitative
p007 / journal page 171 · Results; Metal-phthalocyanines
PbPc and CuPc reproducibility/selectivitypoorer reproducibility and selectivity than MgPcQualitative
Qualitative
p007 / journal page 171 · Results; Metal-phthalocyanines

potentiometric liquid-membrane electrode response

(Ph3Pb)2CO3 o-dichlorobenzene liquid-membrane electrode · Electrode

Na2CO3 solutions at about pH 10; cell potential recorded for carbonate activity at 25 C

Temperature
298
Geometry
Orion series 92 liquid-membrane electrode body with porous membrane support
Context
composite liquid-membrane electrode using organolead salt in o-dichlorobenzene
Measurement source
p005-p006 / journal pages 169-170 · Results; Organolead salts · Fig. 6
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
carbonate lower measurement limitbetween 10^-3 and 10^-4 MrangeText
Range
p010 / journal page 174 · Discussion
carbonate linear activity range10^-1-10^-3 M carbonaterangeText
Range
p005 / journal page 169 · Results; Organolead salts · Fig. 6

potentiometric liquid-membrane electrode response

(Ph3Pb)2SO4 o-dichlorobenzene liquid-membrane electrode · Electrode

Na2SO4 solutions; cell potential recorded for sulphate ion activity with Orion pH meter/printer at 25 C

Temperature
298
Geometry
Orion series 92 liquid-membrane electrode body with porous membrane support
Context
composite liquid-membrane electrode using organolead salt in o-dichlorobenzene
Measurement source
p005-p006 / journal pages 169-170 · Measuring procedure; Results · Fig. 5
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
sulphate electrode interferencehalides, nitrate and hydroxide seriously interferedQualitative
Qualitative
p005 / journal page 169 · Results; Organolead salts
sulphate linear activity range10^-1-10^-5 M sulphaterangeText
Range
p005 / journal page 169 · Results; Organolead salts · Fig. 5
sulphate response slope22 mV decade-1Text
Rounded Reported
p005 / journal page 169 · Results; Organolead salts · Fig. 5