Sensing Application — Rapid and sensitive detection of PD-L1 exosomes using Cu-TCPP 2D MOF as a SPR sensitizer

Measurement evidence

Sensing Application

Rapid and sensitive detection of PD-L1 exosomes using Cu-TCPP 2D MOF as a SPR sensitizer · Wang Y., Mao Z., Chen Q. et al. · Biosensors and Bioelectronics · 2022 · 113954

7 measurement groups · 44 results

Reported values remain attached to the sample, method, conditions, extraction quality and source location that produced them.

Spiked human serum recovery

2D MOF/peptide SPR sensing interface · Thin Film

PD-L1 exosomes spiked into healthy human serum samples; PBS diluted 1000 times; n=3.

Geometry
2D MOF/peptide SPR chip
Context
application sensor in serum matrix
Measurement source
10 · 3.4. Sensitive detection of PD-L1 exosomes · Table 4
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Serum recovery at 5e4 particles/mL101.31%RSD 8.96%Table
Exact Reported
11 · Table 4 · Table 4
Serum recovery at 5e5 particles/mL102.35%RSD 14.60%Table
Exact Reported
11 · Table 4 · Table 4
Serum recovery at 5e6 particles/mL93.43%RSD 5.79%Table
Exact Reported
11 · Table 4 · Table 4
Serum RSD at 5e4 particles/mL8.96%Table
Exact Reported
11 · Table 4 · Table 4
Serum RSD at 5e5 particles/mL14.60%Table
Exact Reported
11 · Table 4 · Table 4
Serum RSD at 5e6 particles/mL5.79%Table
Exact Reported
11 · Table 4 · Table 4

SPR BSA antifouling assay

2D MOF/peptide SPR sensing interface · Thin Film

BSA at 10-500 ng/mL passed through unmodified and peptide-modified 2D MOF chip surfaces.

Geometry
2D MOF and 2D MOF/peptide SPR chips
Context
peptide-modified versus unmodified 2D MOF chip
Measurement source
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 7C-E; Figs. S2-S3
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
BSA response with peptide-modified chipMarked as a best value within this paper0.017 deg at 500 ng/mL BSA17 mdegText
Exact Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 7E
BSA response without peptide modification0.105 deg at 500 ng/mL BSA105 mdegText
Exact Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 7E

SPR PD-L1 exosome calibration and detection limit

2D MOF/peptide SPR sensing interface · Thin Film

PD-L1 exosomes from 1e4 to 5e6 particles/mL flowed through sensor; signal fitted versus log concentration.

Geometry
2D MOF/peptide SPR chip
Context
application sensor
Measurement source
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8C-D
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
PD-L1 exosome calibration upper bound5 x 10^6 particles/mLText
Exact Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8C-D
PD-L1 exosome calibration lower bound10^4 particles/mLText
Exact Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8C-D
Exosome calibration intercept-0.11984 degText
Exact Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8D
Exosome calibration slopeDelta theta = 0.0324 lgCExo - 0.11984Text
Exact Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8D
PD-L1 exosome detection limitMarked as a best value within this paper16.7 particles/mLS/N = 3Text
Exact Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8D
Exosome calibration correlation coefficient R20.989Text
Exact Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8D

SPR peptide/exosome feasibility comparison

2D MOF/peptide SPR sensing interface · Thin Film

Bare gold chip and 2D MOF-chip sequentially exposed to peptides and PD-L1 exosomes under identical conditions.

Geometry
SPR gold chip with peptide-functionalized Cu-TCPP 2D MOF
Context
2D MOF/peptide sensor versus bare gold control
Measurement source
9 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 6C-D
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
PD-L1 exosome SPR angle change on bare goldalmost did not cause a SPR angle changeQualitative
Qualitative
9 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 6D
PD-L1 exosome SPR angle change on MOF sensorMarked as a best value within this paper50.56 mdeg0.05056 degText
Exact Reported
9 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 6D

SPR peptide concentration optimisation

2D MOF/peptide SPR sensing interface · Thin Film

Peptides at 25, 50, 100, 150, and 200 ug/mL flowed over 2D MOF modified gold chip.

Geometry
2D MOF-modified SPR gold chip
Context
composite peptide/MOF sensor
Measurement source
Supplementary information · Fig. S1
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Selected peptide concentrationMarked as a best value within this paper100 ug/mLText
Exact Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 7A-B; Fig. S1

SPR refractive-index sensitivity using NaCl RI standards

2D MOF-chip · Thin Film

RI standards 1.33330-1.36370 flowed over bare gold and 2D MOF-chip; 670 nm MP-SPR instrument.

Geometry
gold SPR chip with 2D MOF sensitizer
Context
2D MOF-chip compared against bare gold control
Measurement source
5 · 3.3. SPR sensitivity analysis · Fig. 5; Tables 1-2
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
RI sensitivity at 20 ug/mL MOF104.77 deg/RIUFigure Axis
Rounded Reported
7 · 3.3. SPR sensitivity analysis · Fig. 5F
RI sensitivity at 40 ug/mL MOF99.50 deg/RIUFigure Axis
Rounded Reported
7 · 3.3. SPR sensitivity analysis · Fig. 5F
RI sensitivity at 5 ug/mL MOF101.33 deg/RIUFigure Axis
Rounded Reported
7 · 3.3. SPR sensitivity analysis · Fig. 5F
RI sensitivity at 7.5 ug/mL MOF129.83 deg/RIUFigure Axis
Rounded Reported
7 · 3.3. SPR sensitivity analysis · Fig. 5F
Absolute RI sensitivity increase after MOF modificationMarked as a best value within this paper39.57 deg/RIUText
Exact Reported
6 · 3.3. SPR sensitivity analysis
Relative RI sensitivity increase after MOF modificationMarked as a best value within this paper40.48%Text
Exact Reported
6 · 3.3. SPR sensitivity analysis
Refractive-index standard range used for SPR sensitivityRI = 1.33330, 1.33500, 1.33675, 1.34170, 1.34986, 1.36370Text
Exact Reported
3 · 2.5. Performance analysis of SPR sensor
Bare gold resonance angle shift3.03 degTable
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 1
Bare gold refractive index change0.03 RIUTable
Rounded Reported
7 · 3.3. SPR sensitivity analysis · Table 1
Bare gold FWHM4.79 degTable
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 1
Bare gold RI sensitivity98.00 deg/RIUTable
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 1
Bare gold Gaussian w4.07 degTable
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 1
MOF-chip resonance angle shiftMarked as a best value within this paper4.30 degTable
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 1
MOF-chip refractive index change0.03 RIUTable
Rounded Reported
7 · 3.3. SPR sensitivity analysis · Table 1
MOF-chip FWHM5.54 degTable
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 1
MOF-chip RI sensitivityMarked as a best value within this paper137.67 deg/RIUTable
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 1
MOF-chip Gaussian w4.71 degTable
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 1
Bare gold detection accuracy0.63Table
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 2
Bare gold figure of merit20.42 RIU-1Table
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 2
MOF-chip detection accuracyMarked as a best value within this paper0.77Table
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 2
MOF-chip figure of meritMarked as a best value within this paper24.81 RIU-1Table
Exact Reported
7 · 3.3. SPR sensitivity analysis · Table 2

SPR selectivity against blood interferents

2D MOF/peptide SPR sensing interface · Thin Film

AA, GSH, glucose, BSA, and PD-L1 exosomes flowed through peptide-functionalized sensor; responses compared.

Geometry
2D MOF/peptide SPR chip
Context
application sensor
Measurement source
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8A-B
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Ascorbic acid SPR response7 mdeg0.007 degText
Rounded Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8B
BSA SPR response15 mdeg0.015 degText
Rounded Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8B
PD-L1 exosome selectivity responseMarked as a best value within this paper97 mdeg0.097 degText
Rounded Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8B
Glucose SPR response8 mdeg0.008 degText
Rounded Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8B
Glutathione SPR response-1 mdeg-0.001 degText
Rounded Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8B
Selectivity significance p-valuep < 0.01<Text
Exact Reported
10 · 3.4. Sensitive detection of PD-L1 exosomes · Fig. 8B