Sensing Application — Sensitive detection of carcinoembryonic antigen (CEA) by a sandwich-type electrochemical immunosensor using MOF-Ce@HA/Ag-HRP-Ab2 as a nanoprobe

Measurement evidence

Sensing Application

Sensitive detection of carcinoembryonic antigen (CEA) by a sandwich-type electrochemical immunosensor using MOF-Ce@HA/Ag-HRP-Ab2 as a nanoprobe · Li W., Ma C., Song Y. et al. · Nanotechnology · 2020 · 185605

6 measurement groups · 37 results

Reported values remain attached to the sample, method, conditions, extraction quality and source location that produced them.

amperometric i-t detection of CEA

sandwich CEA immunosensor on GCE · Electrode

Sandwich-type CEA detection in 20 ml PBS solution at pH 7.0 containing 5 mM H2O2; CV/EIS used 20 ml 5 mM [Fe(CN)6]3-/4- with 0.1 mol l-1 KCl.

Geometry
Au/Ab1/BSA/CEA/MOF-Ce@HA/Ag-HRP-Ab2 modified GCE
Context
final MOF-containing immunosensor
Measurement source
4; 8 · 2.8; 3.2.3 · Figure 3D,E
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
CEA detection limitMarked as a best value within this paper0.2 pg ml-1 (S/N = 3)0.0002 ng ml-1S/N = 3Text
Exact Reported
8 · 3.2.3 · Figure 3D,E
CEA linear response range upper boundMarked as a best value within this paper80 ng ml-180000 pg ml-1Text
Exact Reported
8 · 3.2.3 · Figure 3D,E
CEA linear response range lower boundMarked as a best value within this paper0.001 ng ml-11 pg ml-1Text
Exact Reported
8 · 3.2.3 · Figure 3D,E
CEA calibration regression intercept-38.67 uAText
Exact Reported
8 · 3.2.3 · Figure 3E
CEA calibration R2R2 = 0.9929Text
Exact Reported
8 · 3.2.3 · Figure 3E
CEA calibration regression slope-10.33 uA per lgCText
Exact Reported
8 · 3.2.3 · Figure 3E
CEA test timetested within 30 minText
Rounded Reported
1 · Abstract

literature comparison table

sandwich CEA immunosensor on GCE · Electrode

Analytical performance of the anti-CEA/Ce-MoF@HA/Ag-HRP/GCE sensor compared with other CEA immunosensors.

Geometry
anti-CEA/Ce-MoF@HA/Ag-HRP/GCE
Context
reported best application performance
Measurement source
10 · 3.5 · Table 2
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Table 2 this-work detection limitMarked as a best value within this paper0.0002 ng ml-10.2 pg ml-1Table
Exact Reported
10 · 3.5 · Table 2
Table 2 this-work linear rangeMarked as a best value within this paper0.001-80 ng ml-1Table
Range
10 · 3.5 · Table 2

matrix-effect evaluation against ELISA

sandwich CEA immunosensor on GCE · Electrode

20 fresh serum samples plus 5 quality-control serum samples measured three times by immunosensor and ELISA; logarithmic immunosensor result plotted versus ELISA with 95% prediction interval.

Geometry
final sandwich immunosensor
Context
real-sample application
Measurement source
9-10 · 3.5 · Figure 6
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
matrix-effect samples within prediction interval20 fresh serum samples and 5 serum samples were within the 95% confidence limit95% confidence limitText
Exact Reported
10 · 3.5 · Figure 6

repeatability, selectivity and storage-stability tests

sandwich CEA immunosensor on GCE · Electrode

Five electrodes tested for 5 ng ml-1 CEA; interferences AFP, DA, HRP and BSA at 80 ng ml-1 with 5.0 ng ml-1 CEA; stability checked every five days.

Geometry
final sandwich immunosensor
Context
application validation
Measurement source
8 · 3.4 · Figure 5
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
five-electrode repeatability RSD2.4%five working electrodes; 5 ng ml-1 CEAText
Exact Reported
8 · 3.4 · Figure 5A
interference current responseless than 5.0%upper bound; AFP, DA, HRP and BSA at 80 ng ml-1 with 5.0 ng ml-1 CEAText
Approximate
8 · 3.4 · Figure 5B
current retained after 10 days95.0%Text
Exact Reported
8 · 3.4 · Figure 5C
current retained after 15 days90.2%Text
Exact Reported
8 · 3.4 · Figure 5C
current retained after 20 days83.5%Text
Exact Reported
8 · 3.4 · Figure 5C
current retained after 5 days97.3%Text
Exact Reported
8 · 3.4 · Figure 5C

standard addition recovery in human serum

sandwich CEA immunosensor on GCE · Electrode

CEA additions of 0, 0.1, 1.0, 3.0, 10.0 and 50.0 ng ml-1 to serum samples; immunosensor compared with ELISA, n = 3.

Geometry
final sandwich immunosensor
Context
real-sample application
Measurement source
8-9 · 3.5 · Table 1
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
ELISA serum recovery range96.9%-102.5%n = 3Table
Range
8-9 · 3.5 · Table 1
ELISA serum RSD range2.57% to 4.30%n = 3Table
Range
8-9 · 3.5 · Table 1
ELISA measured CEA for 0 ng ml-1 added spike2.21 +/- 0.06 ng ml-1; recovery 98.2%; RSD 2.71%+/- 0.06 ng ml-1; n = 3Table
Exact Reported
8-9 · 3.5 · Table 1
ELISA measured CEA for 0.1 ng ml-1 added spike2.41 +/- 0.07 ng ml-1; recovery 102.5%; RSD 2.90%+/- 0.07 ng ml-1; n = 3Table
Exact Reported
8-9 · 3.5 · Table 1
ELISA measured CEA for 10 ng ml-1 added spike12.06 +/- 0.31 ng ml-1; recovery 98.4%; RSD 2.57%+/- 0.31 ng ml-1; n = 3Table
Exact Reported
8-9 · 3.5 · Table 1
ELISA measured CEA for 1 ng ml-1 added spike3.33 +/- 0.11 ng ml-1; recovery 99.4%; RSD 3.31%+/- 0.11 ng ml-1; n = 3Table
Exact Reported
8-9 · 3.5 · Table 1
ELISA measured CEA for 3 ng ml-1 added spike5.37 +/- 0.18 ng ml-1; recovery 102.3%; RSD 3.35%+/- 0.18 ng ml-1; n = 3Table
Exact Reported
8-9 · 3.5 · Table 1
ELISA measured CEA for 50 ng ml-1 added spike50.68 +/- 2.18 ng ml-1; recovery 96.9%; RSD 4.30%+/- 2.18 ng ml-1; n = 3Table
Exact Reported
8-9 · 3.5 · Table 1
serum sample CEA concentration for ELISA comparison series2.25 ng ml-1Table
Exact Reported
8-9 · 3.5 · Table 1
serum sample CEA concentration for immunosensor recovery series3.25 ng ml-1Table
Exact Reported
8-9 · 3.5 · Table 1
immunosensor serum recovery range98.6%-104.0%n = 3Text
Range
9 · 3.5 · Table 1
immunosensor serum RSD range2.25% to 3.47%n = 3Text
Range
9 · 3.5 · Table 1
immunosensor measured CEA for 0 ng ml-1 added spike3.28 +/- 0.07 ng ml-1+/- 0.07 ng ml-1; n = 3Table
Exact Reported
8 · 3.5 · Table 1
immunosensor measured CEA for 0.1 ng ml-1 added spike3.37 +/- 0.08 ng ml-1; recovery 100.6%; RSD 2.37%+/- 0.08 ng ml-1; n = 3Table
Exact Reported
8-9 · 3.5 · Table 1
immunosensor measured CEA for 10 ng ml-1 added spike13.06 +/- 0.36 ng ml-1; recovery 98.6%; RSD 2.76%+/- 0.36 ng ml-1; n = 3Table
Exact Reported
8-9 · 3.5 · Table 1
immunosensor measured CEA for 1.0 ng ml-1 added spike4.42 +/- 0.13 ng ml-1+/- 0.13 ng ml-1; n = 3Table
Exact Reported
8 · 3.5 · Table 1
immunosensor measured CEA for 3 ng ml-1 added spike6.34 +/- 0.22 ng ml-1; recovery 101.4%; RSD 3.47%+/- 0.22 ng ml-1; n = 3Table
Exact Reported
8-9 · 3.5 · Table 1
immunosensor measured CEA for 50.0 ng ml-1 added spike54.62 +/- 1.23 ng ml-1+/- 1.23 ng ml-1; n = 3Table
Exact Reported
8 · 3.5 · Table 1

optimisation of PBS pH, antigen incubation time and composite material amount by amperometric current response

sandwich CEA immunosensor on GCE · Electrode

CEA concentration fixed at 10 ng ml-1 while pH, incubation time and material concentration were varied; peak current used as optimisation response.

Geometry
final sandwich immunosensor
Context
application optimisation of the final MOF-containing immunosensor
Measurement source
2-5 · Optimal conditions for immunosensor · Figures S2-S4
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
optimum antigen incubation timeMarked as a best value within this paper40 minsText
Exact Reported
3-4 · Optimal conditions for immunosensor · Figure S3
optimum composite material concentrationMarked as a best value within this paper2 mg/mlText
Exact Reported
3-5 · Optimal conditions for immunosensor · Figure S4
optimum PBS pH for H2O2 reduction processMarked as a best value within this paperpH = 7.0Text
Exact Reported
3-4 · Optimal conditions for immunosensor · Figure S2