DPV calibration for CRP
Au-MOFs-Ab2-CRP/anti-CRP/Pt-COFs/GCE immunosensor · Electrode
Electrochemical response to a series of CRP target concentrations under optimised conditions.
| Property | Reported value | Normalised value | Uncertainty | Origin and quality | Source |
|---|---|---|---|---|---|
| calibration equation intercept | 9.796 x 10^-7 | — | — | Text Exact Reported | 6 · Analytical Performance · Figure 6b |
| calibration equation slopeMarked as a best value within this paper | Delta I_peak = 8.1135 x 10^-6 lg C + 9.796 x 10^-7 | — | — | Text Exact Reported | 6 · Analytical Performance · Figure 6b |
| CRP linear dynamic range upper boundMarked as a best value within this paper | 400 ng/mL | — | — | Text Exact Reported | 6 · Analytical Performance · Figure 6 |
| CRP linear dynamic range lower boundMarked as a best value within this paper | 1 ng/mL | — | — | Text Exact Reported | 6 · Analytical Performance · Figure 6 |
| CRP detection limitMarked as a best value within this paper | 0.2 ng/mL determined by 3delta/slope | — | — | Text Exact Reported | 6 · Analytical Performance · Figure 6 |