ECL calibration for AFB1
AFB1/BSA/Ab/Au NPs/ZnCoMOFs/GCE · Electrode
AFB1 concentration increased from 0.001 ng/mL to 100 ng/mL under optimal conditions
| Property | Reported value | Normalised value | Uncertainty | Origin and quality | Source |
|---|---|---|---|---|---|
| Calibration intercept | 4965.21 | — | — | Text Exact Reported | 6 · 3.6 · Fig. 6B |
| AFB1 detection limitMarked as a best value within this paper | 0.31 pg/mL | — | — | Text Exact Reported | 6 · 3.6 · Fig. 6B |
| Calibration R2 | R2 = 0.99 | — | — | Text Exact Reported | 6 · 3.6 · Fig. 6B |
| AFB1 calibration upper concentration | 100 ng/mL | — | — | Text Exact Reported | 6 · 3.6 · Fig. 6A-B |
| AFB1 calibration lower concentration | 0.001 ng/mL | — | — | Text Exact Reported | 6 · 3.6 · Fig. 6A-B |
| Calibration slope vs lg concentration | -718.68 | — | — | Text Exact Reported | 6 · 3.6 · Fig. 6B |
| Blank ECL intensity mean IB | IB = 7645.8 | — | — | Text Exact Reported | SI text p.S4 · The calculation of detection limit |
| Blank ECL intensity standard deviation SB | SB = 53.1 | — | — | Text Exact Reported | SI text p.S4 · The calculation of detection limit |
| AFB1 detection limit from SI calculation | 309 fg/mL | — | — | Text Exact Reported | SI text p.S4 · The calculation of detection limit |
| Detection-limit signal-to-noise factor k | k = 3 | — | — | Text Exact Reported | SI text p.S4 · The calculation of detection limit |
| Smallest detectable ECL signal IL | IL = 7486.8 | — | — | Text Exact Reported | SI text p.S4 · The calculation of detection limit |