Sensing Application — Target-triggered hybridization chain reaction for ultrasensitive dual-signal miRNA detection

Measurement evidence

Sensing Application

Target-triggered hybridization chain reaction for ultrasensitive dual-signal miRNA detection · Wang Y., Feng H., Huang K. et al. · Biosensors and Bioelectronics · 2022 · 114572

11 measurement groups · 45 results

Reported values remain attached to the sample, method, conditions, extraction quality and source location that produced them.

DPV detection in total RNA extracted from L02 and MCF-7 cells

DNA1-Fc + DNA2-MB/miRNA/MCH/T4-DNA/AuNPs/N-PCD/GCE biosensor · Electrode

Constructed electrochemical sensor used to detect miRNA-21 in total RNA from cell samples.

Atmosphere
N2 for DPV
Geometry
Dual-signal T4-DNA/AuNPs/N-PCD/GCE biosensor
Context
application composite in biological matrix
Measurement source
main p.7 · 3.4 Analytical performance · Fig. 6D
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
L02 cell Fc DPV currentapproximately 1.45 uA from Fig. 6D barvisual estimate from bar heightFigure Axis
Approximate
main p.7 · 3.4 · Fig. 6D
L02 cell MB DPV currentapproximately 0.75 uA from Fig. 6D barvisual estimate from bar heightFigure Axis
Approximate
main p.7 · 3.4 · Fig. 6D
MCF-7 cell Fc DPV currentMarked as a best value within this paperapproximately 3.55 uA from Fig. 6D barvisual estimate from bar heightFigure Axis
Approximate
main p.7 · 3.4 · Fig. 6D
Cell-sample miRNA-21 responseMCF-7 dual-signal peak current significantly higher than L02Text
Qualitative
main p.7 · 3.4 · Fig. 6D
MCF-7 cell MB DPV currentapproximately 3.00 uA from Fig. 6D barvisual estimate from bar heightFigure Axis
Approximate
main p.7 · 3.4 · Fig. 6D

Differential pulse voltammetry (DPV)

DNA1-Fc + DNA2-MB/miRNA/MCH/T4-DNA/AuNPs/N-PCD/GCE biosensor · Electrode

DPV compared final electrode with N-PCD against no-N-PCD Au/GCE control at 2000 fM miRNA-21. SI DPV: 10 mM PBS pH 7.0 under N2, -0.6 to 0.4 V.

Atmosphere
N2 for DPV
Geometry
DNA1-Fc + DNA2-MB/miRNA/MCH/T4-DNA/AuNPs/N-PCD/GCE versus Au/GCE control
Context
application composite with no-N-PCD control
Measurement source
SI p.4 · Electrochemical measurements · Fig. 4D
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Fc DPV response increase from N-PCD147.4%Text
Exact Reported
main p.5 · 3.3 · Fig. 4D
MB DPV response increase from N-PCD200.5%Text
Exact Reported
main p.5 · 3.3 · Fig. 4D

DPV calibration for simultaneous miRNA-21 and miRNA-141

Dual miRNA-21/miRNA-141 biosensor · Electrode

MB signal detects miRNA-21 and Fc signal detects miRNA-141 for 0-10000 fM target concentrations.

Atmosphere
N2 for DPV
Geometry
Dual-target T4-DNA/AuNPs/N-PCD/GCE biosensor
Context
application composite multi-target variant
Measurement source
SI p.7-11 · Analytical performance of biosensors · Fig. S7; Fig. S8
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Multi-target Fc141 calibration equationIFc141 (uA)=1.15 log(c)+17.19Text
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S8B
Multi-target Fc141 correlation coefficient0.998Text
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S8B
Multi-target MB21 calibration equationIMB21 (uA)=0.70 log(c)+10.25Text
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S8B
Multi-target MB21 correlation coefficient0.997Text
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S8B
Multi-target miRNA-141 LOD3.49 fMText
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S8B
Multi-target miRNA-21 LOD6.22 fMText
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S8B

Performance comparison tables

DNA1-Fc + DNA2-MB/miRNA/MCH/T4-DNA/AuNPs/N-PCD/GCE biosensor · Electrode

SI Tables S2 and S3 compare this sensor with literature miRNA-21 sensors and material platforms.

Geometry
T4-DNA/Au/N-PCD+HCR DPV sensor
Context
application composite benchmark
Measurement source
SI p.14-15 · Tables S2 and S3 · Table S2; Table S3
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Table S2 this-work LOD3.74 fMSI Table
Exact Reported
SI p.14 · Table S2 · Table S2
Table S2 this-work linear range5 fM-10000 fMSI Table
Range
SI p.14 · Table S2 · Table S2
Table S3 N-PCD material LOD3.74 fMSI Table
Exact Reported
SI p.15 · Table S3 · Table S3

DPV calibration for miRNA-21

DNA1-Fc + DNA2-MB/miRNA/MCH/T4-DNA/AuNPs/N-PCD/GCE biosensor · Electrode

DPV peak currents of Fc and MB probes recorded for 0, 5, 10, 50, 100, 500, 2000, 5000 and 10000 fM miRNA-21.

Atmosphere
N2 for DPV
Geometry
Dual-signal T4-DNA/AuNPs/N-PCD/GCE biosensor
Context
application composite
Measurement source
main p.6-7 · 3.4 Analytical performance · Fig. 6A,B
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Dual-signal Fc calibration equationIFc (uA) = 1.55 log(c)+23.57Text
Exact Reported
main p.6 · 3.4 · Fig. 6B inset
Dual-signal MB calibration equationIMB (uA) = 1.44 log(c)+21.40Text
Exact Reported
main p.6 · 3.4 · Fig. 6B inset
Dual-signal Fc calibration correlation coefficient0.998Text
Exact Reported
main p.6 · 3.4 · Fig. 6B inset
Dual-signal MB calibration correlation coefficient0.996Text
Exact Reported
main p.6 · 3.4 · Fig. 6B inset
Dual-signal miRNA-21 linear rangeMarked as a best value within this paper5-1.0 x 10^4 fMText
Range
main p.1 · Abstract · Fig. 6B
Dual-signal Fc miRNA-21 LODMarked as a best value within this paper1.92 fMText
Exact Reported
main p.6 · 3.4 · Fig. 6B
Dual-signal MB miRNA-21 LODMarked as a best value within this paper3.74 fMText
Exact Reported
main p.6 · 3.4 · Fig. 6B; Table S2; Table S3

DPV reproducibility and storage stability

DNA1-Fc + DNA2-MB/miRNA/MCH/T4-DNA/AuNPs/N-PCD/GCE biosensor · Electrode

Six electrodes measured for reproducibility; five electrodes stored at 4 deg C for 10 days and tested every 2 days.

Temperature
277
Geometry
Dual-signal T4-DNA/AuNPs/N-PCD/GCE biosensor
Context
application composite
Measurement source
main p.7 · 3.4 Analytical performance · Fig. S2A referenced
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Electrode-to-electrode reproducibility RSD4.9%Text
Exact Reported
main p.7 · 3.4
Fc signal retention after 10 days98.7%Text
Exact Reported
main p.7 · 3.4 · Fig. S2A
MB signal retention after 10 days97.4%Text
Exact Reported
main p.7 · 3.4 · Fig. S2A

DPV selectivity test

DNA1-Fc + DNA2-MB/miRNA/MCH/T4-DNA/AuNPs/N-PCD/GCE biosensor · Electrode

DPV responses to 2000 fM miRNA-21 and interfering RNA sequences.

Atmosphere
N2 for DPV
Geometry
Dual-signal T4-DNA/AuNPs/N-PCD/GCE biosensor
Context
application composite
Measurement source
main p.6-7 · 3.4 Analytical performance · Fig. 6C; Table S1
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Selectivity interferent concentration2000 fMText
Exact Reported
main p.6 · 3.4 · Fig. 6C
Selectivity against interfering RNAsno significant difference from absence of targetText
Qualitative
main p.6 · 3.4 · Fig. 6C

Serum recovery experiment

DNA1-Fc + DNA2-MB/miRNA/MCH/T4-DNA/AuNPs/N-PCD/GCE biosensor · Electrode

Serum samples from healthy subjects and breast cancer patients centrifuged and diluted 10 times with 10 mM PBS pH 7.4; recoveries summarised in Table S4.

Atmosphere
N2 for DPV
Geometry
Dual-signal T4-DNA/AuNPs/N-PCD/GCE biosensor
Context
application composite in serum matrix
Measurement source
main p.7 · 3.4 Analytical performance · Table S4 referenced
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Maximum serum recovery106.57%SI Table
Exact Reported
SI p.16 · Table S4 · Table S4
Minimum serum recovery92.16%SI Table
Exact Reported
SI p.16 · Table S4 · Table S4
Serum recovery range92.16% to 106.57%Text
Range
main p.7 · 3.4 · Table S4 referenced
Serum sample 3 endogenous miRNA-21 found33.5 +/- 1.7 fM+/- 1.7 fMSI Table
Exact Reported
SI p.16 · Table S4 · Table S4
Serum sample 4 endogenous miRNA-21 found36.9 +/- 2.4 fM+/- 2.4 fMSI Table
Exact Reported
SI p.16 · Table S4 · Table S4

DPV optimisation of sensor construction variables

DNA1-Fc + DNA2-MB/miRNA/MCH/T4-DNA/AuNPs/N-PCD/GCE biosensor · Electrode

Optimised Au deposition time, miRNA incubation time, DNA1-Fc/DNA2-MB concentration and co-incubation time using 2000 fM miRNA.

Geometry
Dual-signal T4-DNA/AuNPs/N-PCD/GCE biosensor
Context
application composite
Measurement source
SI p.5-8 · Optimization of sensor · Fig. S1
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Selected Au deposition time30 sText
Exact Reported
SI p.5 · Optimization of sensor · Fig. S1A
Optimal Au deposition time by Fc signal25 sText
Exact Reported
SI p.5 · Optimization of sensor · Fig. S1A
Optimal/selected Au deposition time by MB signal30 sText
Exact Reported
SI p.5 · Optimization of sensor · Fig. S1A
Selected miRNA incubation time70 minText
Exact Reported
SI p.5 · Optimization of sensor · Fig. S1B
Selected DNA1-Fc/DNA2-MB co-incubation time70 minText
Exact Reported
SI p.6 · Optimization of sensor · Fig. S1D
Selected DNA1-Fc/DNA2-MB concentration1.0 mMText
Exact Reported
SI p.5-6 · Optimization of sensor · Fig. S1C

DPV calibration for miRNA-21 using Fc single signal

Single-signal Fc miRNA-21 biosensor · Electrode

Fc DPV signal for 0, 5, 10, 50, 100, 500, 2000, 5000 and 10000 fM miRNA-21.

Atmosphere
N2 for DPV
Geometry
Single-signal Fc T4-DNA/AuNPs/N-PCD/GCE biosensor
Context
application composite single-signal comparator
Measurement source
SI p.6-9 · Analytical performance of biosensors · Fig. S3; Fig. S4
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Single-signal Fc calibration equationIFc (uA)=1.58 log(c)+23.94Text
Exact Reported
SI p.6-7 · Analytical performance of biosensors · Fig. S4B
Single-signal Fc miRNA-21 LOD2.08 fMText
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S4B
Single-signal Fc calibration correlation coefficient0.995Text
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S4B

DPV calibration for miRNA-21 using MB single signal

Single-signal MB miRNA-21 biosensor · Electrode

MB DPV signal for 0, 5, 10, 50, 100, 500, 2000, 5000 and 10000 fM miRNA-21.

Atmosphere
N2 for DPV
Geometry
Single-signal MB T4-DNA/AuNPs/N-PCD/GCE biosensor
Context
application composite single-signal comparator
Measurement source
SI p.7-10 · Analytical performance of biosensors · Fig. S5; Fig. S6
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Single-signal MB calibration equationIMB (uA)=1.63 log(c)+24.20Text
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S6B
Single-signal MB miRNA-21 LOD3.89 fMText
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S6B
Single-signal MB calibration correlation coefficient0.995Text
Exact Reported
SI p.7 · Analytical performance of biosensors · Fig. S6B