chemiresistive gas sensing
Cu3(HITP)2 NFA chemiresistor sensor · Electrode
Room-temperature response-recovery to ethanol vapour at 1, 5, 10, 20 and 50 ppm; response defined as delta R/R0 in high-purity dry air.
| Property | Reported value | Normalised value | Uncertainty | Origin and quality | Source |
|---|---|---|---|---|---|
| ethanol response linear-fit R2 | R2 = 0.96 | — | — | Caption Exact Reported | 11 · Figure caption · Figure 6c |
| theoretical ethanol limit of detectionMarked as a best value within this paper | about 0.5 ppm | — | about | Text Approximate | 12 · 3.2 Chemiresistive Gas Sensors · Figure 6c |
| recovery time for 20 ppm ethanol | tau_rec = 64 s | — | — | Text Exact Reported | 11 · 3.2 Chemiresistive Gas Sensors · Figure 6b |
| ethanol response at 10 ppm | approximately 2.45% | — | visual estimate from Figure 6a/c | Figure Axis Approximate | 11 · 3.2 Chemiresistive Gas Sensors · Figure 6a,c |
| ethanol response at 1 ppm | approximately 1.4% | — | visual estimate from Figure 6a/c | Figure Axis Approximate | 11 · 3.2 Chemiresistive Gas Sensors · Figure 6a,c |
| ethanol response at 20 ppm | approximately 3.3% | — | visual estimate from Figure 6a/c | Figure Axis Approximate | 11 · 3.2 Chemiresistive Gas Sensors · Figure 6a,c |
| ethanol response at 50 ppmMarked as a best value within this paper | approximately 5.1% | — | visual estimate from Figure 6a/c | Figure Axis Approximate | 11 · 3.2 Chemiresistive Gas Sensors · Figure 6a,c |
| ethanol response at 5 ppm | approximately 1.9% | — | visual estimate from Figure 6a/c | Figure Axis Approximate | 11 · 3.2 Chemiresistive Gas Sensors · Figure 6a,c |
| response time for 20 ppm ethanol | tau_res = 50 s | — | — | Text Exact Reported | 11 · 3.2 Chemiresistive Gas Sensors · Figure 6b |
| chemiresistor nickel lead wire diameter | 50 micrometre diameter nickel wire | — | — | Text Exact Reported | S2 · Gas sensor characterizations |