Electrical Transport — Lithium Ion Diffusion in a Metal-Organic Framework Mediated by an Ionic Liquid

Measurement evidence

Electrical Transport

Lithium Ion Diffusion in a Metal-Organic Framework Mediated by an Ionic Liquid · Fujie K., Ikeda R., Otsubo K. et al. · Chemistry of Materials · 2015 · 7355-7361

10 measurement groups · 18 results

Reported values remain attached to the sample, method, conditions, extraction quality and source location that produced them.

AC impedance, quasi-four-probe method

ELT-bulk · Unknown

Solartron 1260 impedance/gain-phase analyser and 1296 dielectric interface; 1 Hz to 1 MHz; liquid in Kapton washer between stainless steel electrodes; cooled below 210 K then measured stepwise during heating.

Temperature
below 210 K to above 350 K
Geometry
Kapton washer approximately 0.1 mm thick x 3.1 mm inner diameter; stainless steel electrodes
Context
bulk lithium-doped ionic liquid control
Measurement source
2 · Experimental Section
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
ELT-bulk conductivity inflection temperaturearound 248 KText
Approximate
4 · Results and Discussion · Figure 8
ELT-bulk log ionic conductivity near 300 Kvisual estimate from Figure 8 at ca. 300 K: log10(sigma / S cm-1) approximately -3.20.00063 S cm^-1approximately +/-0.3 log unitsFigure Axis
Approximate
4 · Results and Discussion · Figure 8

AC impedance, quasi-four-probe method

ELT@Z100 · Pellet

Powder pressed into pellet with carbon sheet electrodes on both faces; 1 Hz to 1 MHz; cooled below 210 K then measured stepwise during heating.

Temperature
below 210 K to above 350 K
Geometry
pellet approximately 0.5 mm thick x 3.0 mm diameter; carbon sheet electrodes
Context
guest-loaded ZIF-8
Measurement source
2 · Experimental Section
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Bulk IL vs 100% occupied IL@ZIF-8 conductivityET-bulk and ELT-bulk are 2 orders of magnitude higher than ET@Z100 and ELT@Z100Text
Approximate
5 · Results and Discussion · Figure 8
ELT@Z100 log ionic conductivity near 300 Kvisual estimate from Figure 8 at ca. 300 K: log10(sigma / S cm-1) approximately -5.70.000002 S cm^-1approximately +/-0.3 log unitsFigure Axis
Approximate
4 · Results and Discussion · Figure 8
Effect of LiTFSA addition on IL@ZIF-8 conductivityELT@Z100 and ELT@Z75 showed lower conductivities and larger inclinations than ET@Z100 and ET@Z75Text
Qualitative
4 · Results and Discussion · Figure 8
ELT@Z100 conductivity low-temperature dropNo marked decrease was detected even at low temperatureText
Qualitative
4 · Results and Discussion · Figure 8
100% occupancy vs 75% occupancy conductivityMarked as a best value within this paperET@Z100 and ELT@Z100 are 2 orders of magnitude higher than ET@Z75 and ELT@Z75Text
Approximate
5 · Results and Discussion · Figure 8

AC impedance, quasi-four-probe method

ELT@Z75 · Pellet

Powder pressed into pellet with carbon sheet electrodes on both faces; 1 Hz to 1 MHz; cooled below 210 K then measured stepwise during heating.

Temperature
below 210 K to above 350 K
Geometry
pellet approximately 0.5 mm thick x 3.0 mm diameter; carbon sheet electrodes
Context
guest-loaded ZIF-8
Measurement source
2 · Experimental Section
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
ELT@Z75 log ionic conductivity near 300 Kvisual estimate from Figure 8 at ca. 300 K: log10(sigma / S cm-1) approximately -8.26.3e-9 S cm^-1approximately +/-0.4 log unitsFigure Axis
Approximate
4 · Results and Discussion · Figure 8

AC impedance, quasi-four-probe method

ET-bulk · Unknown

Solartron 1260 impedance/gain-phase analyser and 1296 dielectric interface; 1 Hz to 1 MHz; liquid in Kapton washer between stainless steel electrodes; cooled below 210 K then measured stepwise during heating.

Temperature
below 210 K to above 350 K
Geometry
Kapton washer approximately 0.1 mm thick x 3.1 mm inner diameter; stainless steel electrodes
Context
bulk ionic liquid control
Measurement source
2 · Experimental Section
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
ET-bulk conductivity inflection temperaturearound 265 KText
Approximate
4 · Results and Discussion · Figure 8

AC impedance, quasi-four-probe method

ET@Z100 · Pellet

Powder pressed into pellet with carbon sheet electrodes on both faces; 1 Hz to 1 MHz; cooled below 210 K then measured stepwise during heating.

Temperature
below 210 K to above 350 K
Geometry
pellet approximately 0.5 mm thick x 3.0 mm diameter; carbon sheet electrodes
Context
guest-loaded ZIF-8
Measurement source
2 · Experimental Section
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
ET@Z100 log ionic conductivity near 300 Kvisual estimate from Figure 8 at ca. 300 K: log10(sigma / S cm-1) approximately -4.90.000013 S cm^-1approximately +/-0.3 log unitsFigure Axis
Approximate
4 · Results and Discussion · Figure 8

AC impedance, quasi-four-probe method

ET@Z75 · Pellet

Powder pressed into pellet with carbon sheet electrodes on both faces; 1 Hz to 1 MHz; cooled below 210 K then measured stepwise during heating.

Temperature
below 210 K to above 350 K
Geometry
pellet approximately 0.5 mm thick x 3.0 mm diameter; carbon sheet electrodes
Context
guest-loaded ZIF-8
Measurement source
2 · Experimental Section
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource

Nyquist impedance plots from AC impedance measurements

ELT@Z100 · Pellet

SI Figure S7 gives ELT@Z100 Nyquist plots during heating at 241, 268, 295, and 343 K.

Temperature
241, 268, 295, 343
Geometry
pellet approximately 0.5 mm thick x 3.0 mm diameter; carbon sheet electrodes
Context
guest-loaded ZIF-8
Measurement source
S6 · Supporting Information · Figure S7
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
ELT@Z100 Nyquist plot temperatures241, 268, 295, and 343 KCaption
Exact Reported
S6 · Supporting Information · Figure S7

Nyquist impedance plots from AC impedance measurements

ELT@Z75 · Pellet

SI Figure S6 gives ELT@Z75 Nyquist plots during heating at 282, 296, 309, and 344 K.

Temperature
282, 296, 309, 344
Geometry
pellet approximately 0.5 mm thick x 3.0 mm diameter; carbon sheet electrodes
Context
guest-loaded ZIF-8
Measurement source
S5 · Supporting Information · Figure S6
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
ELT@Z75 Nyquist plot temperatures282, 296, 309, and 344 KCaption
Exact Reported
S5 · Supporting Information · Figure S6

7Li pulsed-field gradient NMR (PFG-NMR) diffusion measurement

ELT-bulk · Unknown

7Li at 155.5 MHz using Bruker Diff 50 probe; hermetically sealed glass tubes; stimulated-echo pulse sequence; g = 0 to 2500 G cm^-1, Delta = 40 ms, delta = 1 ms.

Temperature
approximately 400
Atmosphere
sealed glass tube
Context
bulk lithium-doped ionic liquid control
Measurement source
2 · Experimental Section
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
Activation energy for lithium self-diffusion in ELT-bulk18.6 kJ mol-1Text
Exact Reported
5 · Results and Discussion · Figure 10
ELT-bulk log lithium self-diffusion coefficient near 400 Kvisual estimate from Figure 10 near 400 K: log10(DLi / m2 s-1) approximately -9.751.8e-10 m^2 s^-1approximately +/-0.2 log unitsFigure Axis
Approximate
5 · Results and Discussion · Figure 10

7Li pulsed-field gradient NMR (PFG-NMR) diffusion measurement

ELT@Z100 · Pellet

7Li at 155.5 MHz using Bruker Diff 50 probe; hermetically sealed glass tubes; stimulated-echo pulse sequence; g = 0 to 2500 G cm^-1, Delta = 40 ms, delta = 1 ms.

Temperature
approximately 403
Atmosphere
sealed glass tube
Context
guest-loaded ZIF-8
Measurement source
2 · Experimental Section
PropertyReported valueNormalised valueUncertaintyOrigin and qualitySource
ELT@Z100 lithium self-diffusion coefficient relative to bulkDLi of ELT@Z100 was 2 orders of magnitude lower than ELT-bulkText
Approximate
5 · Results and Discussion · Figure 10
Activation energy for lithium self-diffusion in ELT@Z100Marked as a best value within this paper15.7 kJ mol-1Text
Exact Reported
5 · Results and Discussion · Figure 10
ELT@Z100 log lithium self-diffusion coefficient near 403 Kvisual estimate from Figure 10 near 403 K: log10(DLi / m2 s-1) approximately -11.354.5e-12 m^2 s^-1approximately +/-0.2 log unitsFigure Axis
Approximate
5 · Results and Discussion · Figure 10
PFG-NMR signal dependencestimulated-echo signal decreased with PFG magnitude, conforming to the Stejskal-Tanner equation in both ELT-bulk and ELT@Z100Text
Qualitative
5 · Results and Discussion · Figure 9