DPV clinical serum comparison
Apt/Au/Cu-THQ/GCE · Electrode
Human serum samples from sepsis cases measured by Apt/Au/Cu-THQ/GCE and compared with WB-80 microbial dynamic monitoring system.
| Property | Reported value | Normalised value | Uncertainty | Origin and quality | Source |
|---|---|---|---|---|---|
| Clinical serum sample 1 LPS by biosensorMarked as a best value within this paper | 1.33 +/- 0.29 pg/mL; WB-80 <5.00 pg/mL | 1.33 pg/mL | +/- 0.29 pg/mL; n = 3 | Table Exact Reported | 9 · Serum sample detection · Table 2 |
| Clinical serum sample 2 LPS by biosensor | 8.47 +/- 1.07 pg/mL; WB-80 8.08 pg/mL | 8.47 pg/mL | +/- 1.07 pg/mL; n = 3 | Table Exact Reported | 9 · Serum sample detection · Table 2 |
| Clinical serum sample 3 LPS by biosensor | 15.15 +/- 1.35 pg/mL; WB-80 15.43 pg/mL | 15.15 pg/mL | +/- 1.35 pg/mL; n = 3 | Table Exact Reported | 9 · Serum sample detection · Table 2 |
| WB-80 standard method LOD | LOD = 5 pg/mL | 5 pg/mL | — | Text Exact Reported | 9 · Serum sample detection · Table 2 |