differential pulse voltammetry (DPV)
Cu3(HHTP)2/SPE · Electrode
0.1 mol/L PBS; DPV range -1 to 1 V; potential increment 0.005 V; pulse amplitude 0.05 V; pulse width 0.3 s; AA concentration range 25-1645 umol/L.
| Property | Reported value | Normalised value | Uncertainty | Origin and quality | Source |
|---|---|---|---|---|---|
| AA calibration R2 | R2 = 0.99 | — | — | Text Exact Reported | 5 · 2. Results and Discussion · Figure 4b |
| AA detection linear range upper boundMarked as a best value within this paper | 1645 umol/L | 1645 umol/L | — | Text Exact Reported | 5 · 2. Results and Discussion · Figure 4a,b |
| AA detection linear range lower boundMarked as a best value within this paper | 25 umol/L | 25 umol/L | — | Text Exact Reported | 5 · 2. Results and Discussion · Figure 4a,b |
| AA calibration slope | y = -0.0345x - 2 x 10-6 | — | — | Text Exact Reported | 5 · 2. Results and Discussion · Figure 4b |
| AA detection limitMarked as a best value within this paper | 2.44 umol/L (S/N = 3) | 2.44 umol/L | S/N = 3 | Text Exact Reported | 8 · 4. Conclusions · Figure 4a,b |